Evaporative Water Loss
During metabolic measurements at temperatures above freezing, evaporative water loss was determined gravimetrically. Upstream from the chemical columns, an aliquot of air (0.1 L/min) was drawn off the exhaust line and diverted for a timed interval through a series of preweighed (0.1 mg)
-tubes containing Drierite®. The aliquot then passed through a second series of
-tubes containing Ascarite® before entering the oxygen analysis system. Evaporative water loss was calculated using [Eq. 1]
| Ė = (mw·.Ve)/(.Va·t·m) | Eq. 1 |
where Ė is evaporative water loss (mg·g-1·h-1), mw is mass of water collected (mg), .Ve is rate of air flow into the chamber (3.0 L/min), .Va is the rate of air flow through the