Testing the cable.—1. See that the testing key is open and the shunt at 0. Connect the earth lead to ground on the cable armor. Remove the earth connection from No. 1 conductor and connect the cable lead to this conductor; in wet weather the connector joint should be dipped in melted paraffin. (In using paraffin to insulate joints or ends bring it just above 212° F. to evaporate any moisture present. It should not be boiling. The paraffin coating should be at least as thick as the rubber insulation and extend back over the rubber for an inch or more.)

2. Close the testing key to the left (+ to earth), stopping at the discharge position, and turn the shunt to the unity post. There should be no deflection. If there is, it is due either to a charge on the cable, which will disappear after a moment, or to earth currents. (It is assumed that the testing apparatus has been thoroughly tested for insulation.) If due to earth currents, the conductor is probably a poor one. Earth currents are readily recognizable by their fluctuating character. Before assuming that the trouble can not be removed, the joint between the lead and the conductor should be examined again. Moisture on the cable end will give a path for earth currents. Note the value and direction of the throw of the galvanometer and record it.

3. Turn the shunt to 0, close the testing key all the way down (+ to earth), noting the time to the second, or starting the stop watch at the same time, if one is available. The time must be accurately noted. The insulation resistance at the end of one minute’s electrification is the resistance to be reported.

4. When 35 seconds have elapsed, turn the shunt to the ¹/₁₀₀₀-post and watch the galvanometer throw; if small, move the shunt successively to the ¹/₁₀₀-post, to the ¹/₁₀-post, and to the unity post. This operation must be completed before 45 seconds have elapsed from the time the key was closed. With good cable the unity post will always be reached without danger of throwing the galvanometer reading off the scale. Remember that each successive post should give 10 times the throw of the preceding post.

5. At the end of one minute read the deflection, correct for the leakage of the leads and the earth currents, and record. (See example following.)

6. At the end of two minutes read the deflection, correct and record it. For good cable it should be less than the deflection observed at the end of one minute.

7. Turn the shunt to 0, and reverse the key, stopping at the discharge position. Turn the shunt on gradually until the unity post is reached and wait until the reading is 0, indicating that the conductor is discharged. If earth currents are present, 0 will not be reached or will be passed. In this case proceed as before described. A submarine mine cable conductor a mile long will discharge ordinarily in about three minutes.

8. Turn the shunt to 0, stop and start the stop watch; at the same time close the key all the way down (- to earth).

9. After 35 seconds, start turning the shunt, ceasing at 45 seconds. ([See paragraph 4, above].)

10. At the end of one minute read the deflection, correct and record it. For good cable it should be substantially the same as the deflection observed at the end of one minute with + of the battery to earth.