If a new rider is purchased, or an old one replaced, care must be taken that its weight corresponds to the graduations on the beam of the balance on which it is to be used. The weight of the rider in milligrams must be equal to the number of large divisions (5, 6, 10, or 12) between the central knife-edge and the knife-edge at the end of the beam. It should be noted that on some balances the last division bears no number. Each new rider should be tested against a 5 or 10-milligram weight.

In some of the most recent forms of the balance a chain device replaces the smaller weights and the use of the rider as just described.

Before using a balance, it is always best to test its adjustment. This is absolutely necessary if the balance is used by several workers; it is always a wise precaution under any conditions. For this purpose, brush off the balance pans with a soft camel's hair brush. Then note (1) whether the balance is level; (2) that the mechanism for raising and lowering the beams works smoothly; (3) that the pan-arrests touch the pans when the beam is lowered; and (4) that the needle swings equal distances on either side of the zero-point when set in motion without any load on the pans. If the latter condition is not fulfilled, the balance should be adjusted by means of the adjusting screw at the end of the beam unless the variation is not more than one division on the scale; it is often better to make a proper allowance for this small zero error than to disturb the balance by an attempt at correction. Unless a student thoroughly understands the construction of a balance he should never attempt to make adjustments, but should apply to the instructor in charge.

The object to be weighed should be placed on the left-hand balance pan and the weights upon the right-hand pan. Every substance which could attack the metal of the balance pan should be weighed upon a watch-glass, and all objects must be dry and cold. A warm body gives rise to air currents which vitiate the accuracy of the weighing.

The weights should be applied in the order in which they occur in the weight-box (not at haphazard), beginning with the largest weight which is apparently required. After a weight has been placed upon the pan the beam should be lowered upon its knife-edges, and, if necessary, the pan-arrests depressed. The movement of the pointer will then indicate whether the weight applied is too great or too small. When the weight has been ascertained, by the successive addition of small weights, to the nearest 5 or 10 milligrams, the weighing is completed by the use of the rider. The correct weight is that which causes the pointer to swing an equal number of divisions to the right and left of the zero-point, when the pointer traverses not less than five divisions on either side.

The balance case should always be closed during the final weighing, while the rider is being used, to protect the pans from the effect of air currents.

Before the final determination of an exact weight the beam should always be lifted from the knife-edges and again lowered into place, as it frequently happens that the scale pans are, in spite of the pan-arrests, slightly twisted by the impact of the weights, the beam being thereby virtually lengthened or shortened. Lifting the beam restores the proper alignment.

The beam should never be set in motion by lowering it forcibly upon the knife-edges, nor by touching the pans, but rather by lifting the rider (unless the balance be provided with some of the newer devices for the purpose), and the swing should be arrested only when the needle approaches zero on the scale, otherwise the knife-edges become dull. For the same reason the beam should never be left upon its knife-edges, nor should weights be removed from or placed on the pans without supporting the beam, except in the case of the small fractional weights.

When the process of weighing has been completed, the weight should be recorded in the notebook by first noting the vacant spaces in the weight-box, and then checking the weight by again noting the weights as they are removed from the pan. This practice will often detect and avoid errors. It is obvious that the weights should always be returned to their proper places in the box, and be handled only with pincers.

It should be borne in mind that if the mechanism of a balance is deranged or if any substance is spilled upon the pans or in the balance case, the damage should be reported at once. In many instances serious harm can be averted by prompt action when delay might ruin the balance.