GRADUATION OF FLASKS
It is a general custom to purchase the flasks ungraduated and to graduate them for use under standard conditions selected for the laboratory in question. They may be graduated for "contents" or "delivery." When graduated for "contents" they contain a specified volume when filled to the graduation at a specified temperature, and require to be washed out in order to remove all of the solution from the flask. Flasks graduated for "delivery" will deliver the specified volume of a liquid without rinsing. A flask may, of course, be graduated for both contents and delivery by placing two graduation marks upon it.
PROCEDURE.—To calibrate a flask for !contents!, proceed as follows: Clean the flask, using a chromic acid solution, and dry it carefully outside and inside. Tare it accurately; pour water into the flask until the weight of the latter counterbalances weights on the opposite pan which equal in grams the number of cubic centimeters of water which the flask is to contain. Remove any excess of water with the aid of filter paper (Note 1). Take the flask from the balance, stopper it, place it in a bath at the desired temperature, usually 15.5° or 17.5°C., and after an hour mark on the neck with a diamond the location of the lowest point of the meniscus (Note 2). The mark may be etched upon the flask by hydrofluoric acid, or by the use of an etching ink now commonly sold on the market.
To graduate a flask which is designed to !deliver! a specified volume, proceed as follows: Clean the flask as usual and wipe all moisture from the outside. Fill it with distilled water. Pour out the water and allow the water to drain from the flask for three minutes. Counterbalance the flask with weights to the nearest centigram. Add weights corresponding in grams to the volume desired, and add distilled water to counterbalance these weights. An excess of water, or water adhering to the neck of the flask, may be removed by means of a strip of clean filter paper. Stopper the flask, place it in a bath at 15.5°C. or 17.5°C. and, after an hour, mark the location of the lowest point of the meniscus, as described above.
[Note 1: The allowable error in counterbalancing the water and weights varies with the volume of the flask. It should not exceed one ten-thousandth of the weight of water.]
[Note 2: Other methods are employed which involve the use of calibrated apparatus from which the desired volume of water may be run into the dry flask and the position of the meniscus marked directly upon it. For a description of a procedure which is most convenient when many flasks are to be calibrated, the student is referred to the !Am. Chem J.!, 16, 479.]
GENERAL DIRECTIONS FOR VOLUMETRIC ANALYSES
It cannot be too strongly emphasized that for the success of analyses uniformity of practice must prevail throughout all volumetric work with respect to those factors which can influence the accuracy of the measurement of liquids. For example, whatever conditions are imposed during the calibration of a burette, pipette, or flask (notably the time allowed for draining), must also prevail whenever the flask or burette is used.
The student should also be constantly watchful to insure parallel conditions during both standardization and analyst with respect to the final volume of liquid in which a titration takes place. The value of a standard solution is only accurate under the conditions which prevailed when it was standardized. It is plain that the standard solutions must be scrupulously protected from concentration or dilution, after their value has been established. Accordingly, great care must be taken to thoroughly rinse out all burettes, flasks, etc., with the solutions which they are to contain, in order to remove all traces of water or other liquid which could act as a diluent. It is best to wash out a burette at least three times with small portions of a solution, allowing each to run out through the tip before assuming that the burette is in a condition to be filled and used. It is, of course, possible to dry measuring instruments in a hot closet, but this is tedious and unnecessary.
To the same end, all solutions should be kept stoppered and away from direct sunlight or heat. The bottles should be shaken before use to collect any liquid which may have distilled from the solution and condensed on the sides.